Objectives:
Ureteral double-J stents are known to migrate proximally and distally within the urinary tract, while perforation and stent displacement are uncommon. Possible mechanisms of displacement are either original malpositioning with ureteral perforation or subsequent fistula and erosion of the excretory system, due to infection or long permanence of the device. We present the unique case of complete intraperitoneal stent migration in a 59-year-old caucasian male without evidence of urinary fistula at the moment of diagnosis, so far an unreported complication.
Materials And Methods:
Eight months after the placement of a double-J stent for lower right ureteral stricture at a district hospital, the patient came at our observation for urosepsis and hydro-uretero-nephrosis. A CT scan demonstrated intraperitoneal migration of the stent outside the urinary tract. Cystoscopy failed to visualize the lower extremity of the stent, a percutaneous nephrostomy was placed to drain the urinary system and the stent was removed through a small abdominal incision on the right lower quadrant.
Results:
In our case we presume that during the positioning manoeuvre the guide wire perforated simultaneously the lower ureteral wall and the pelvic peritoneum, and that once the upper end of the stent was coiled, the lower extremity was also attracted intraperitoneally. The lack of pain due to the spinal lesion concurred to this unusual complication.
Conclusions:
We must be aware that ureteral double J stents may be found displaced even inside the peritoneal cavity, and that the use of retrograde pyelography during placement is of paramount importance to exclude misplacement of an apparently normally coiled upper extremity of the stent.
More Related Videos
10:09Application of Simplified Stent-bridging Pancreaticogastrostomy in Open Pancreaticoduodenectomy
Published on: March 17, 2026
07:48Intraoperative Video Consultation Following Bile Duct Transection Facilitates Direct OR Transfer for Robotic Hepaticojejunostomy at Tertiary Center
Published on: January 9, 2026
Related Concept Videos
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Extracorporeal Removal of Drugs: Peritoneal Dialysis and Hemodialysis
Inflammatory Bowel Disease V: Surgical Management
Here are some common surgical interventions for IBD:
Esophageal Perforation-I: Introduction
The location of esophageal perforation can vary, occurring anywhere along the esophagus....