Factors influencing delivered mean airway pressure during nasal CPAP with the RAM cannula
Jeffrey S Gerdes1,2,3, Emidio M Sivieri1,2, Soraya Abbasi1,2,3
1Section on Newborn Pediatrics, Pennsylvania Hospital, Philadelphia, Pennsylvania.
Objective:
To measure mean airway pressure (MAP) delivered through the RAM Cannula® when used with a ventilator in CPAP mode as a function of percent nares occlusion in a simulated nasal interface/test lung model and to compare the results to MAPs using a nasal continuous positive airway pressure (NCPAP) interface with nares fully occluded.
Study Design:
An artificial airway model was connected to a spontaneous breathing lung model in which MAP was measured at set NCPAP levels between 4 and 8 cmH2 O provided by a Dräger Evita XL® ventilator and delivered through three sizes of RAM cannulae. Measurements were performed with varying leakage at the nasal interface by decreasing occlusion from 100% to 29%, half-way prong insertion, and simulated mouth leakage. Comparison measurements were made using the Dräger BabyFlow® NCPAP interface with a full nasal seal.
Results:
With simulated mouth closed, the Dräger interface delivered MAPs within 0.5 cmH2 O of set CPAP levels. For the RAM cannula, with 60-80% nares occlusion, overall delivered MAPs were 60 ± 17% less than set CPAP levels (P < 0.001). Further, MAP decreased progressively with decreasing percent nares occlusion. The simulated open mouth condition resulted in significantly lower MAPs to <1.7 cmH2 O. The one-half prong insertion depth condition, with closed mouth, yielded MAPs approximately 35 ± 9% less than full insertion pressures (P < 0.001).
Conclusions:
In our bench tests, the RAM interface connected to a ventilator in NCPAP mode failed to deliver set CPAP levels when applied using the manufacturer recommended 60-80% nares occlusion, even with closed mouth and full nasal prong insertion conditions.
More Related Videos
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation I: Indication and Settings
Cardiopulmonary Resuscitation II: ACLS Airway Management
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...