Related Concept Videos

Imaging Studies for Cardiovascular System III: X-Ray01:20

Imaging Studies for Cardiovascular System III: X-Ray

684
The most common cardiovascular diagnostic test is an X-ray. It produces images of the heart, blood vessels, and adjacent structures.
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
684

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Can the current radiation dose, in chest tomosynthesis, be reduced with retained image quality? A study in the context of lung cancer screening programs.

PloS one·2026
Same author

Survey of radiological optimization processes in Swedish hospitals-similarities and differences between different modalities.

Radiation protection dosimetry·2026
Same author

Assessing the influence of kernel selection on chest computed tomography image quality across varying dose levels using TrueFidelity reconstruction.

Radiation protection dosimetry·2026
Same author

Optimisation in X-ray and molecular imaging 2025-editorial.

Radiation protection dosimetry·2026
Same author

From digital chest tomosynthesis to 3D CT.

Radiation protection dosimetry·2026
Same author

Modelling the relative precision of whole-kidney dosimetry in molecular radiotherapy using a power law approach.

Radiation protection dosimetry·2026

Related Experiment Video

Updated: Apr 15, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
08:30

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging

Published on: September 11, 2011

15.0K

Estimating effective dose from 3D imaging with interventional fluoroscopy systems using limited exposure data.

Angelica Svalkvist1, Jonny Hansson1, Magnus Båth2

  • 1Department of Radiation Physics, Institute of Clinical Sciences, Sahlgrenska Academy at University of Gothenburg, SE-413 45 Gothenburg, Sweden Department of Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, SE-413 45 Gothenburg, Sweden.

Acta Radiologica (Stockholm, Sweden : 1987)
|April 9, 2015
PubMed
Summary
This summary is machine-generated.

Simplified dose calculations for 3D rotational imaging using average exposure values provide reasonably accurate effective dose estimations, with errors generally under 20%. This method is suitable for interventional fluoroscopy systems.

Keywords:
Effective dosedose-area productinterventional radiologyradiation protectionthree-dimensional (3D) imaging

More Related Videos

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
06:20

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition

Published on: March 11, 2021

7.9K
Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
07:57

Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform

Published on: March 24, 2022

3.3K

Related Experiment Videos

Last Updated: Apr 15, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
08:30

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging

Published on: September 11, 2011

15.0K
Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
06:20

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition

Published on: March 11, 2021

7.9K
Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
07:57

Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform

Published on: March 24, 2022

3.3K

Area of Science:

  • Medical Physics
  • Radiological Imaging
  • Radiation Dosimetry

Background:

  • Effective dose estimations in 3D rotational imaging are challenging.
  • Interventional fluoroscopy systems often lack individual projection exposure values.

Purpose of the Study:

  • To quantify the error in effective dose calculations.
  • Investigate the impact of using average vs. individual exposure values in 3D rotational imaging.

Main Methods:

  • Compared effective dose calculations using individual exposure values versus mean values with distributed dose-area product (DAP).
  • Utilized anthropomorphic phantoms and simulated worst-case scenarios for interventional fluoroscopy.
  • Assessed errors across varying anatomical regions, tube voltages, and patient sizes.

Main Results:

  • Simplifying calculations with average exposure values introduced errors within ±14%.
  • Worst-case scenarios showed errors below ±20% for effective dose estimations.
  • Errors were consistent across different patient sizes and anatomical regions.

Conclusions:

  • Dose calculations using average exposure values offer reasonably accurate estimations for 3D imaging.
  • This simplified method is practical for interventional fluoroscopy systems despite inherent uncertainties.
  • The findings support using average exposure values for radiation dose assessments.