Related Experiment Video
Updated: Apr 15, 2026

Multi-analyte Biochip MAB Based on All-solid-state Ion-selective Electrodes ASSISE for Physiological Research
Published on: April 18, 2013
Modified screen-printed ion selective electrodes for potentiometric determination of sodium dodecylsulfate in
Tamer Awad Ali1, Gehad G Mohamed
1Egyptian Petroleum Research Institute (EPRI), 11727, Cairo, Egypt.
Abstract:
Fabrication and general performance characteristics of novel screen-printed sensors for potentiometric determination of sodium dodecylsulfate (SDS) are described. The sensors are based on the use of ion-association complexes of SDS with cetylpyridinium chloride (electrode I) and cetyltrimethylammonium bromide (electrode II) as exchange sites in a screen-printed electrode matrix. Electrodes (I) and (II) show fast, stable, and near-Nernstian response for the mono-charge anion of SDS over the concentration range of 1×10(-2) - 5.8×10(-7) and 1×10(-2)-6.3×10(-7) mol/L at 25°C and the pH range of 2.0-9.0 and 2.0-8.0 with anionic slope of 57.32±0.81 and 56.58±0.65 mV/decade, respectively. Electrodes (I) and (II) have lower LODs of 5.8×10(-7) and 6.3×10(-7) mol/L and response times of about 8 and 13 s, respectively. Shelf life of 5 months for both electrodes is adequate. Selectivity coefficients of SDS related to a number of interfering cations, and some inorganic compounds were investigated. There were negligible interferences caused by most of the investigated species. The direct determination of 0.10-13.50 mg of SDS by electrodes (I) and (II) shows average recoveries of 99.96 and 99.85%, and mean RSDs of 0.83 and 1.04%, respectively. In the present investigation, both electrodes were used successfully as end point indicators for determination of SDS in pure pharmaceutical preparations and real spiked water samples. The results obtained using the proposed sensors to determine SDS in solution compared favorably with those obtained by the standard addition method.
Related Concept Videos
Potentiometry: Membrane Electrodes
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
SDS-PAGE
A variation of gel electrophoresis, termed polyacrylamide gel electrophoresis (PAGE), is commonly used for separating proteins according to their molecular size by passing them through a polyacrylamide gel. Because of the varying charges associated with amino acid side chains, PAGE can be used to separate intact...
Potentiometry: Overview
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential...