Prospective study on dosimetric comparison of helical tomotherapy and 3DCRT for craniospinal irradiation - A single
Anna Bandurska-Luque1, Tomasz Piotrowski2, Agnieszka Skrobała2
1Radiotherapy Department II, Greater Poland Cancer Centre, Poznań, Poland.
Aim:
This prospective study aims to assess feasibility of helical tomotherapy (HT) for craniospinal irradiation (CSI) and perform dosimetric comparison of treatment plans for both HT and 3D conformal radiotherapy (3DCRT).
Background:
CSI is a challenging procedure. Large PTV size requires field matching due to technical limitations of standard linear accelerators, which cannot irradiate such volumes as a single field. HT could help to avoid these limitations as irradiation of long fields is possible without field matching.
Materials And Methods:
Three adults were enrolled from 2009 to 2010. All patients received radiochemotherapy. Treatment plans in prone position for 3DCRT and in supine position for HT were generated. The superior plan was used for patients' irradiation. Plans were compared with the application of DVH, Dx parameters - where x represents a percentage of the structure volume receiving a normalized dose and homogeneity index (HI).
Results:
All patients received HT irradiation. The treatment was well tolerated. The HT plans resulted in a better dose coverage and uniformity in the PTV: HI were 5.4, 7.8, 6.8 for HT vs. 10.3, 6.6, 10.4 for 3DCRT. For most organs at risk (OARs), the D(V80) was higher for HT than for 3DCRT, whereas D(V5) was lower for HT.
Conclusions:
HT is feasible for CSI, and in comparison with 3DCRT it improves PTV coverage. HT reduces high dose volumes of OARs, but larger volumes of normal tissue receive low radiation dose. HT requires further study to establish correlations between dosimetrical findings and clinical outcomes, especially with regard to late sequelae of treatment.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
11:24Stem Cell Transplantation Strategies for the Restoration of Cognitive Dysfunction Caused by Cranial Radiotherapy
Published on: October 18, 2011
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography