A New Method to Measure the Subpubic Arch Angle Using 3-D Ultrasound
Tullio Ghi1, Aly Youssef, Federica Martelli
1Department of Obstetrics and Gynecology, Sant'Orsola-Malpighi University Hospital, University of Bologna, Bologna, Italy.
Fetal Diagnosis and Therapy
|April 15, 2015
Summary
The Oblique View eXtended Imaging (OVIX) technique reliably measures the subpubic angle (SPA) in pregnant women. The 3-D multiplanar technique (MPT) overestimates SPA measurements compared to OVIX.
Area of Science:
- Obstetrics and Gynecology
- Medical Imaging
- Fetal Biometry
Background:
- Accurate measurement of the subpubic angle (SPA) is crucial for assessing pelvic dimensions in pregnant women.
- Evaluating new imaging techniques for SPA measurement is essential for improving obstetric assessments.
Purpose of the Study:
- To assess the reproducibility of the Oblique View eXtended Imaging (OVIX) and 3-D multiplanar technique (MPT) for subpubic angle measurement.
- To evaluate the intermethod agreement between OVIX and MPT in term gestation women.
Main Methods:
- Transperineal 3-D ultrasound volumes were acquired from 155 women at term gestation.
- Subpubic angles were measured twice by one operator and once by another using both OVIX and MPT.
- Reproducibility and intermethod agreement were analyzed using intraclass correlation coefficients (ICC) and Bland-Altman methods.
Main Results:
- OVIX demonstrated high intraobserver (ICC 0.912) and good interobserver (ICC 0.791) agreement for SPA measurement.
- MPT showed moderate intraobserver (ICC 0.573) and good interobserver (ICC 0.640) agreement.
- Good intermethod agreement was found between OVIX and MPT (ICC 0.614), though MPT measurements were significantly wider.
Conclusions:
- The Oblique View eXtended Imaging (OVIX) technique is a reliable method for subpubic angle measurement.
- The 3-D multiplanar technique (MPT) tends to overestimate subpubic angle measurements compared to OVIX.
- Further research is warranted to determine the clinical utility of these techniques.