Abstract:
Hematocrit (HCT)-based assay bias (composed of area and recovery bias) is an important contributing factor to the barriers that currently hinder the development and acceptance of dried blood spots (DBS) as a widely used quantitative bioanalytical sampling technique for regulatory studies. This article describes the evaluation of a practical internal standard spray addition technique, used prior to LC-MS/MS analysis, which is demonstrated to nullify the effect of recovery bias. To our knowledge, this is the first time a potential solution to HCT-based recovery bias has been investigated in detail and reported in the literature. This new technique is coupled with accurate volume DBS sampling, whole-spot extraction, and automated direct elution techniques to demonstrate a workflow that both nullifies HCT-based assay bias and the additional manual extraction burden associated with DBS analysis.