Challenges in standardization of blood pressure measurement at the population level
Hanna Tolonen1, Päivikki Koponen2, Androniki Naska3
1Department of Health, National Institute for Health and Welfare, Helsinki, Finland. hanna.tolonen@thl.fi.
Background:
Accurate blood pressure measurements are needed in clinical practice, intervention studies and health examination surveys. Blood pressure measurements are sensitive: their accuracy can be affected by measurement environment, behaviour of the subject, measurement procedures, devices used for the measurement and the observer. To minimize errors in blood pressure measurement, a standardized measurement protocol is needed.
Methods:
The European Health Examination Survey (EHES) Pilot project was conducted in 2009-2012. A pilot health examination survey was conducted in 12 countries using a standardized protocol. The measurement protocols used in each survey, training provided for the measurers, measurement data, and observations during site visits were collected and evaluated to assess the level of standardization.
Results:
The EHES measurement protocol for blood pressure was followed accurately in all 12 pilot surveys. Most of the surveys succeeded in organizing a quiet and comfortable measurement environment, and staff instructed survey participants appropriately before examination visits. In all surveys, blood pressure was measured three times, from the right arm in a sitting posture. The biggest variation was in the device used for the blood pressure measurement.
Conclusions:
It is possible to reach a high level of standardization for blood pressure measurements across countries and over time. A detailed, standardized measurement protocol, and adequate training and monitoring during the fieldwork and centrally organized quality assessment of the data are needed. The recent EU regulation banning the sale of mercury sphygmomanometer in European Union Member States has set new challenges for the standardization of measurement devices since the validity of oscillometric measurements is device-specific and performance of aneroid devices depends very much on calibration.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure