Related Experiment Videos
Argon laser ablation of malignant ventricular tachycardia associated with coronary artery disease
S Saksena1, I Gielchinsky, N G Tullo
1Division of Cardiology, Newark Beth Israel Medical Center, New Jersey.
Abstract:
The long-term clinical efficacy and safety of intraoperative mapping-guided argon laser ablation alone or in conjunction with standard surgical methods were assessed in 20 consecutive patients with refractory sustained ventricular tachycardia (VT) or ventricular fibrillation. A 15-W argon ion gas laser was used and pulsed laser energy was delivered through a fiberoptic catheter delivery system. Pre- and intraoperative mapping was used to localize the arrhythmogenic tissue. Postoperative clinical, ambulatory electrocardiographic and electrophysiologic evaluations were performed before discharge and at 1 year of follow-up. Thirty-eight VT morphologies were mapped and ablated with laser energy alone (82%), combined laser ablation and mechanical resection (13%) or mechanical resection alone (5%). Concomitant coronary artery bypass surgery was performed in 15 patients and in 1 patient it was performed with mitral value replacement. Postoperative 30-day mortality was 5%. One patient (5%) required postoperative antiarrhythmic drug therapy, and all survivors had suppression of inducible sustained VT at discharge. Mean left ventricular ejection fraction increased from 34 +/- 12% preoperatively to 41 +/- 13% postoperatively (p = 0.001). Efficacy rates for ablation of VT sites associated with anterior myocardial infarction and inferior or posterior myocardial infarction were comparable (100 vs 96%, respectively, p greater than 0.2). At 1-year follow-up no sudden deaths had occurred and total survival rate was 90%. Intraoperative pulsed argon laser ablation alone or in conjunction with standard surgical techniques improves the efficacy of surgical ablation procedures for VT or ventricular fibrillation and reduces the need for additional postoperative antiarrhythmic drug or device therapy.
Insights
Argon laser ablation guided by intraoperative mapping effectively treats refractory ventricular tachycardia (VT) and ventricular fibrillation (VF), improving patient outcomes and reducing the need for further therapies.
Area of Science:
- Cardiovascular Surgery
- Electrophysiology
- Medical Technology
Background:
- Refractory sustained ventricular tachycardia (VT) and ventricular fibrillation (VF) pose significant clinical challenges.
- Current surgical ablation methods may have limitations in efficacy and require adjunctive therapies.
Purpose of the Study:
- To evaluate the long-term clinical efficacy and safety of intraoperative mapping-guided argon laser ablation for VT/VF.
- To assess the effectiveness of laser ablation alone or combined with standard surgical techniques.
Main Methods:
- Twenty patients with refractory VT/VF underwent ablation using a 15-W argon ion laser delivered via a fiberoptic catheter.
- Pre- and intraoperative mapping guided the localization and ablation of arrhythmogenic tissue.
- Patients were followed clinically and with electrocardiographic and electrophysiologic evaluations for 1 year.
Main Results:
- Thirty-eight VT morphologies were ablated, primarily with laser energy alone (82%).
- Postoperative 30-day mortality was 5%; 90% overall survival at 1 year.
- All survivors were free of inducible VT at discharge, with improved left ventricular ejection fraction (34% to 41%, p=0.001).
Conclusions:
- Intraoperative mapping-guided argon laser ablation is an effective and safe treatment for refractory VT/VF.
- This technique improves surgical ablation efficacy and reduces the need for postoperative antiarrhythmic drugs or devices.
- Comparable efficacy was observed for ablating VT sites related to anterior and inferior/posterior myocardial infarction.