Development of a novel straining holder for transmission electron microscopy compatible with single tilt-axis
K Sato1, H Miyazaki2, T Gondo2
1Institute for Materials Research, Tohoku University, 2-1-1 Katahira, Aoba-ku, Sendai 980-8577, Japan ksato@imr.tohoku.ac.jp.
Abstract:
We have developed a newly designed straining specimen holder for in situ transmission electron microscopy (TEM) compatible with high-angle single tilt-axis electron tomography. The holder can deform a TEM specimen under tensile stress with the strain rate between 1.5 × 10(-6) and 5.2 × 10(-3) s(-1). We have also confirmed that the maximum tilt angle of the specimen holder reaches ±60° with a rectangular shape aluminum specimen. The new specimen holder, termed as 'straining and tomography holder', will have wide range potential applications in materials science.