Ventilator gas delivery wave form substantially impacts plateau pressure and peak-to-plateau pressure gradient
Adrian Anthony Maung1, Dirk C Johnson, Gina M Luckianow
1From the Department of Surgery (A.A.M., D.C.J., G.M.L.), Section of Trauma, Surgical Critical Care and Surgical Emergencies, Yale School of Medicine, New Haven, Connecticut; and Division of Trauma, Surgical Critical Care and Emergency Surgery (L.J.K.), Perelman School of Medicine, University of Pennsylvania; and Department of Surgery (L.J.K.), Philadelphia Veterans Affairs Medical Center, Philadelphia, Pennsylvania.
Background:
To determine whether plateau pressure (Pplat) measurement is lowered and peak airway pressure (Pawpeak)-to-Plat gradient is increased by measurement on a decelerating compared with square gas delivery wave form.
Methods:
Prospective before and after study of mechanically ventilated injured and critically ill patients in an adult surgical intensive care unit. Pplat, Pawpeak, and Pawpeak-to-Pplat gradient were measured on decelerating and square gas delivery wave forms.
Results:
Pplat and other routine ventilator parameters were measured in 82 (47 trauma, 35 emergency general surgery) consecutive convenience sampled adult intensive care unit patients on decelerating and then square gas delivery wave forms. Peak gas flow was fixed at 40 L/min; all other parameters (rate, tidal volume, positive end-expiratory pressure) were held constant. All patients were managed on assist control volume cycled ventilation using fentanyl and midazolam or propofol; no neuromuscular blockade was used. Patients with Pawpeak more than 35 cm H2O were excluded. Comparing decelerating with square gas delivery, mean Pawpeak was lower (25.1 ± 2.3 cm H2O vs. 33.1 ± 2.1 cm H2O; p < 0.0001) and mean Pplat was lower (21.3 ± 1.9 cm H2O vs. 24.8 ± 2.5 cm H2O; p < 0.0001), resulting in a decreased Pawpeak-to-Pplat gradient (3.8 ± 2.1 vs. 8.3 ± 2.3; p < 0.0001).
Conclusion:
Changing from a decelerating to a square gas delivery wave form significantly increases Pplat and Pawpeak, thereby increasing the Pawpeak-to-Pplat gradient. This increase may prompt unwarranted therapy aimed at reducing the gradient to its normal value of 4 cm H2O pressure or less. Conversely, patients with a high Pawpeak on a square wave form may benefit from transitioning to a decelerating wave form before changing ventilation parameters.
Level Of Evidence:
Diagnostic study, level III.
More Related Videos
06:22Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome ARDS
Published on: April 7, 2021
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...