Sensorless Viscosity Measurement in a Magnetically-Levitated Rotary Blood Pump
Wataru Hijikata1, Jun Rao2, Shodai Abe2
1Precision and Intelligence Laboratory, Tokyo Institute of Technology, Yokohama, Japan.
Artificial Organs
|April 30, 2015
Summary
This study introduces a novel sensorless method to measure blood viscosity in implantable rotary blood pumps. This innovation enables more accurate flow rate control by assessing viscosity without extra components.
Area of Science:
- Biomedical Engineering
- Fluid Dynamics
- Medical Devices
Background:
- Accurate flow rate control in implantable rotary blood pumps is crucial for matching physiological demand.
- Existing methods for flow rate estimation often require flow meters or lack precise blood viscosity measurements.
- Noninvasive, sensorless methods for blood viscosity are needed for advanced pump control.
Purpose of the Study:
- To develop and validate a sensorless method for measuring blood viscosity in magnetically levitated rotary blood pumps.
- To enable accurate blood viscosity assessment without additional sensors or mechanical components.
- To explore the potential of this method for improving flow rate estimation in blood pumps.
Main Methods:
- A sensorless viscosity measurement technique was developed for magnetically levitated rotary blood pumps.
- Vibrational excitation was applied to the impeller using magnetic bearings.
- Viscosity was determined by measuring the phase difference between magnetic bearing current and impeller displacement.
Main Results:
- The developed method achieved high correlation (R² > 0.992) with reference viscosity measurements.
- Mean absolute deviation for measured viscosity was 0.12 mPa·s across a range of 1.18 to 5.12 mPa·s.
- The technique demonstrated reliable performance for various working fluids.
Conclusions:
- A novel, sensorless method for measuring blood viscosity in rotary blood pumps has been successfully developed.
- This technique eliminates the need for additional sensors, simplifying pump design and improving reliability.
- The validated method shows significant potential for enhancing the accuracy of flow rate control in implantable blood pumps.
Related Concept Videos
Measurement of Fluid Pressure
2.2K
Fluid pressure is commonly measured using devices called manometers, which rely on liquid columns to indicate pressure differences. The height of a liquid column in a manometer reflects the pressure exerted by the fluid, providing a simple yet effective means of measurement. Different types of manometers serve specific purposes based on their configurations and the type of fluids involved.
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
2.2K
Measurement of Blood Pressure
4.2K
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
4.2K