Related Experiment Video
Updated: Apr 13, 2026

Three-dimensional Optical-resolution Photoacoustic Microscopy
Published on: May 3, 2011
Rapid and noncontact photoacoustic tomography imaging system using an interferometer with high-speed phase modulation
Jun Liu1, Zhilie Tang1, Yongbo Wu1
1School of Physics and Telecom Engineering, South China Normal University, Guangzhou 510006, China.
Abstract:
We designed, fabricated, and tested a rapid and noncontact photoacoustic tomography (PAT) imaging system using a low-coherence interferometer with high-speed phase modulation technique. Such a rapid and noncontact probing system can greatly decrease the time of imaging. The proposed PAT imaging system is experimentally verified by capturing images of a simulated tissue sample and the blood vessels within the ear flap of a mouse (pinna) in vivo. The axial and lateral resolutions of the system are evaluated at 45 and ∼15 μm, respectively. The imaging depth of the system is 1 mm in a special phantom. Our results show that the proposed system opens a promising way to realize noncontact, real-time PAT.