Computational Modeling of Neonatal Cardiopulmonary Bypass Hemodynamics With Full Circle of Willis Anatomy

Senol Piskin1, Akif Ündar2, Kerem Pekkan1,3

  • 1Department of Mechanical Engineering, Koc University, Istanbul, Turkey.

Artificial Organs
|May 6, 2015
PubMed

Insights

Accurate head-neck blood perfusion during neonatal cardiopulmonary bypass (CPB) requires including the 3D Circle of Willis (CoW). Cannula placement significantly impacts blood flow distribution, crucial for surgical strategy and minimizing complications.

Area of Science:

  • Biomedical Engineering
  • Computational Fluid Dynamics
  • Pediatric Cardiology

Background:

  • Cardiopulmonary bypass (CPB) is vital for congenital heart defect (CHD) repair in neonates.
  • Cannula placement in CPB critically influences hemodynamic performance.
  • Previous computational studies simplified the neonatal cerebral system.

Purpose of the Study:

  • To integrate a 3D patient-specific cerebral model, including the Circle of Willis (CoW), with a hypoplastic neonatal aortic arch for enhanced CPB simulations.
  • To compare hemodynamic differences between innominate artery (IA) and ductus arteriosus (DA) cannulation configurations.
  • To assess the impact of CoW inclusion on predicting head-neck blood perfusion.

Main Methods:

  • Developed a 3D patient-specific computational model integrating MRI-derived cerebral vasculature (including CoW) with a neonatal aortic arch.
  • Utilized a commercial, experimentally validated CFD solver for flow simulations at Re=2150.
  • Applied steady velocity inlet boundary conditions for the CPB pump and resistance boundary conditions for all outlets.

Main Results:

  • Inclusion of the 3D CoW is essential for accurate head-neck blood perfusion prediction.
  • Significant differences in flow distribution were observed between IA- and DA-cannula configurations.
  • DA-cannulation resulted in higher flow to the ascending aorta (198%) and cerebral arteries (91% right, 68% left) compared to IA-cannulation. Retrograde vertebral artery flow occurred with IA-cannulation but not DA-cannulation.

Conclusions:

  • The integrated 3D CoW model is critical for preoperative planning of neonatal CPB cannulation strategies.
  • Cannula configuration significantly alters blood flow dynamics, impacting perfusion to vital organs.
  • Findings support improved neonatal arterial cannula design to mitigate CPB-related cerebral complications.