Endotracheal Tube Seal and Suction Performance in a Novel Biorealistic Tracheal Model
Sarah K Rozycki1, Fredrick P Dixon2, Melissa A Yopp1
1Department of Pediatrics, Virginia Commonwealth University School of Medicine and the Children's Hospital of Richmond at Virginia Commonwealth University, Richmond, Virginia.
Background:
Endotracheal tube (ETT) cuffs create a seal to protect against secretion entry to the lungs. Cuff inflation currently is recommended at 20-30 cm H2O pressure. ETT designs have variable seal performance in bench studies using rigid tracheal models lacking the dynamic characteristics of the human trachea. We compared ETT designs within a new, biorealistic tracheal model to assess cuff and suction performance in the setting of a compliant trachea.
Methods:
Three ETT designs (Mallinckrodt Hi-Lo, KimVent Microcuff, and Sheridan/HVT) were tested for performance by simulant leakage below the cuff and air leakage (measured as return tidal volume ≥ 80% delivered) over a range of cuff (5-25 cm H2O) and end-expiratory pressure (PEEP 0-15 cm H2O). Subglottic suction channel performance was tested in 2 ETTs (TaperGuard Evac [Covidien] and ISIS HVT [Teleflex]) as time to evacuate the simulant.
Results:
All ETT cuffs provided effective seals at an inflation pressure of 12 cm H2O when PEEP was ≤ 5 cm H2O. The Microcuff ETT sealed at the lowest pressure of 6 cm H2O, whereas the Sheridan/HVT cuff sealed at 12 cm H2O (P = .01). With a PEEP of 15 cm H2O, a reciprocal increase in air leak occurred, requiring a cuff inflation up to 22 cm H2O to maintain a return tidal volume at ≥ 80% delivered. Suction channel performance improved in the lateral position compared with supine for both ETT designs during continuous 15 mm Hg suction pressure (P = .001).
Conclusions:
Within a novel model with normal trachea compliance, we found all ETT designs tested to seal at lower than current recommended cuff pressures.
More Related Videos
09:10Heterotopic and Orthotopic Tracheal Transplantation in Mice used as Models to Study the Development of Obliterative Airway Disease
Published on: January 20, 2010
04:43Endotracheal Intubation via Tracheotomy and Subsequent Thoracotomy in Rats for Non-Survival Applications
Published on: March 15, 2024
Related Concept Videos
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Tracheostomy Suctioning II: Procedure
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...