Related Experiment Video
Updated: Apr 12, 2026

Multi-Scale Modification of Metallic Implants With Pore Gradients, Polyelectrolytes and Their Indirect Monitoring In vivo
Published on: July 1, 2013
Detailed inspection of metal implants
Harry S Hothi1, Reshid Berber, Robert K Whittaker
11 Institute of Orthopaedics and Musculoskeletal Science, Stanmore - UK.
Abstract:
Detailed visual inspection of metal hips is the first step in retrieval analysis. In this study a systematic visual inspection protocol was developed to quantify bearing surface changes and their associations with material loss was investigated. Simple and multiple linear regression models found that moderate surface scratching, discolouration, haziness and the size of visible wear scars were all significantly associated with material loss (R2 = 5%-73%, p<0.05). Visual inspection is not a substitute for measurement of material loss but an understanding of bearing surface changes may offer unique clues as to the mechanisms of failure of retrieved hips.