Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
Radiographic distortion artifact of circular external fixators
Scott Secrest1, Jesse Nagy, Stephen Kneller
1From the Department of Veterinary Biosciences and Diagnostic Imaging, University of Georgia, Athens, GA (S.S.); IDEXX Imaging Services, Columbia, MO (J.N.); and Department of Veterinary Clinical Medicine, College of Veterinary Medicine, University of Illinois, Mahomet, IL (S.K.).
Circular external fixators can cause radiographic distortion. This study found distortion is greatest at smaller wire angles and when further from the X-ray table, potentially impacting patient care.
Area of Science:
- Orthopedic surgery
- Medical imaging
- Radiography
Background:
- Circular external fixators are vital orthopedic devices.
- Radiographic distortion can misrepresent transfixing wire location, affecting patient outcomes.
- Understanding distortion is crucial for accurate interpretation.
Purpose of the Study:
- To evaluate radiographic distortion in three circular external fixator systems.
- To determine the impact of transfixing wire angles (30°, 60°, 90°) on distortion.
- To identify optimal positioning to minimize artifact.
Main Methods:
- Assessed distortion of three circular external fixator apparatuses.
- Varied transfixing wire angles relative to the X-ray beam (30°, 60°, 90°).
- Measured distortion at different degrees of rotation from parallel to the central X-ray beam.
Main Results:
- Maximum distortion occurred at 10° rotation from parallel for all apparatuses.
- Distortion was most significant with transfixing wires at a 30° angle.
- Smaller angles and greater distance from the X-ray table increased distortion.
Conclusions:
- Distortion artifact is most severe at smaller wire angles and increased distance from the table.
- Positioning the fixator parallel to the X-ray beam and close to the table minimizes artifact.
- Proper positioning aids accurate radiographic interpretation and prevents unnecessary procedures.