Background:
Access port problems after laparoscopic adjustable gastric banding can be significant complications. The Easyband™ is an innovative type of gastric band, whose internal diameter can be adjusted by a telemetrically activated motor. The aim of this study was to evaluate safety, efficacy and performance of the Easyband™.
Methods:
A total of 110 morbidly obese patients were prospectively enrolled. The Easyband™ was implanted laparoscopically by experienced bariatric surgeons in six different hospital locations. Patient characteristics, surgery details, postoperative weight loss, and complications were recorded. Follow-up lasted 2 years.
Results:
Follow-up was completed by 78.2 % of the patients. Surgeons rated 71 % of each aspect of Easyband™ implantation as "easy" or "very easy". Adjustments were successful in 91 % of the attempts. One or more adverse events occurred in 79.1 % of the patients. Thirty-six serious adverse events were reported, of which 50 % was device-related. One or more functional tests failed in 20.9 % of the devices, 8.2 % passed after a repeated test. A reintervention was necessary in 15.5 % of the patients. Mean weight loss after 2 years was 24.2 ± 14.0 kg, mean excess weight loss was 46.1 ± 24 %.
Conclusion:
Excess weight loss with the Easyband™ was comparable with other gastric banding devices. Adjustment of the device was simple, non-invasive and more acceptable to patients than with a standard access port. However, a high incidence of device-related problems requiring surgical explantation occurred. The Easyband™ represents a major advance in gastric band design, but significant technical problems need to be resolved before further implants can be recommended.
More Related Videos
Related Concept Videos
Dose-Response Relationship: Potency and Efficacy
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...