An Analysis of 3 Common CardioGen-82 82Rb Infusion System Injection Methods and Their Impact on Clinical Volume and
Amanda E Roby1, Nils P Johnson1, K Lance Gould2
1Weatherhead PET Center for Preventing and Reversing Atherosclerosis, University of Texas Medical School at Houston, Houston, Texas.
Unlabelled:
In the wake of the Food and Drug Administration (FDA) recall, many clinics have had to reduce their examination volumes to meet the new generator volume usage requirements. This review tests 3 common infusion methods and how they affect patient dose, generator volume usage, image counts, and generator volume limits.
Methods:
Three common configurations of the (82)Rb infusion system settings--standard 50-mL, volume-limiting, and bolus methods--were tested to determine how they affect patient dose, generator volume, and image counts. Each injection configuration was tested daily for the duration of 3 consecutive generators by injection into separate vials. Each injection configuration was also infused into a beaker and imaged to determine the impact of image counts for each method. The total estimated volumes for multiple examination and quality assurance clinical situations were simulated to observe the use of each method relative to the new FDA volume alert and expiration limits.
Results:
Vial tests confirmed that the bolus method used the least amount of volume per infusion and stayed the most consistent throughout the life of the generator. The bolus method also produced a lower patient dose after approximately 10 d of use. The beaker tests in the scanner showed that the standard 50-mL method produced the greatest number of total counts for the flow and uptake images. On the basis of the estimated total volume simulations, the bolus method allowed for the most examinations over the life of the generator while staying within the new FDA limits.
Conclusion:
All 3 methods for augmenting the (82)Rb infusion system produced different outcomes for patient dose, image counts, and total generator volume use. The standard 50-mL method ensured the maximum amount of counts available for imaging throughout the life of the generator. The bolus method provided a consistent and predictable amount of volume use. The volume-limiting method fell somewhere in the middle of volume predictability and count preservation.
More Related Videos
07:13An Alternative and Validated Injection Method for Accessing the Subretinal Space via a Transcleral Posterior Approach
Published on: December 7, 2016
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Drug Accumulation During Multiple Dosing: Repetitive IV Injections
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Imaging Studies VII: Vascular Imaging
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...