Related Experiment Video
Updated: Apr 12, 2026

Implantation of Electrospun Vascular Grafts with Optimized Structure in a Rat Model
Published on: June 27, 2018
Electron microscopy study of a vascular prosthesis destructed in vivo reveals fractures in Dacron fibers
Witold Woźniak1, Wojciech Olszewski2, Grzegorz Górski3
1Department of General and Vascular Surgery, Faculty of Medicine with the English Division and the Physiotherapy Division, Medical University of Warsaw, Poland drww@onet.pl.
Abstract:
The genuine destruction of a synthetic prosthesis wall, as a late complication of vascular surgery, is extremely rare. We report a case of a 64-year-old male who had his 12-year-old femoropopliteal synthetic graft explanted due to two large pseudoaneurysms in the middle section of the graft. Microscopic evaluation demonstrated the areas of focal thinning along the entire prosthesis wall, with "foreign body" type reaction in the adjacent connective tissue. Transmission electron microscopy showed longitudinal fractures of Dacron fibers interposed with cellular structures, suggesting that destruction must have taken place significantly earlier. The problems of limited graft durability and graft surveillance are discussed.