Quantitative Analysis of Chloramphenicol in Royal Jelly by Column-switching LC-MS/MS Using a Pretreatment Column with
Shin-ichi Kawano1, Yoshihiro Hayakawa, Yuki Hashi
1Shimadzu Global COE for Application & Technical Development, Shimadzu (China) Co., Ltd.
Abstract:
An on-line pretreatment liquid chromatography-tandem mass spectrometry (LC-MS/MS) system was developed for the analysis of chloramphenicol (CAP) in royal jelly. A novel methylcellulose-immobilized restricted access media column with a higher-pressure capability of 60 MPa (MC-ODS HP) was developed for the effective removal of proteins and other compounds in the sample matrix. CAP in a sample solution was extracted in 2 min by the column-switching LC-MS/MS system. The system provides a minimum sample pretreatment along with highly sensitive and reproducible analysis. As a result, the limit of quantitation of CAP was 10 pg/mL (= 0.1 μg/kg royal jelly) and the linear dynamic range was between 10 and 10000 pg/mL (correlation coefficient greater than 0.999). The proposed method meets the requirements of regulations in EU (0.3 μg/kg). The inter-day precision and accuracy of CAP at 100 pg/mL over 3 days were 4.5 and 95.4%, respectively. Compared with the conventional method with a pressure of below 25 MPa, the peak separation in the MRM chromatogram was improved by using smaller particles (1.6 μm) for the analytical ODS column. The LC-MS/MS system with an MC-ODS HP expanded the applicability of the automated pretreatment.
More Related Videos
08:56Quantitative Analysis of the Cellular Lipidome of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: March 8, 2020
08:53Quantitative and Qualitative Method for Sphingomyelin by LC-MS Using Two Stable Isotopically Labeled Sphingomyelin Species
Published on: May 7, 2018
Related Concept Videos
Principles Of Column Chromatography
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...