Fluctuation of Cup Orientation During Press-Fit Insertion: A Possible Cause of Malpositioning
Takashi Nishii1, Takashi Sakai2, Masaki Takao2
1Department of Orthopaedic Medical Engineering, Osaka University Graduate School of Medicine, Suita, Osaka, Japan.
Abstract:
We evaluated real-time angular changes during press-fit cup insertion in 70 total hip arthroplasties. Inclination and anteversion cup angles on navigation system were recorded after each procedure of five hammer blows during cup insertion, and maximum deviated inclination (MDI) and anteversion (MDA) angles until the cup was fully seated were measured. We found 1 (1%) and 29 (41%) hips with increases and decreases in MDI of 5° or more, and 13 (19%) and 6 (8%) hips with increases and decreases in MDA of 5° or more, respectively. The MDI in men was significantly lower and the MDA for left-side surgery was significantly higher. Surgeons should pay attention to fluctuation in angular changes during cup insertion and may consider countermeasures to minimize cup malpositioning.
Related Concept Videos
Unsymmetric Bending - Angle of Neutral Axis
When a bending moment is applied at an angle θ concerning the vertical axis of a symmetrical member, it can be resolved into components along the member's principal...
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Plastic Deformation in Circular Shafts
Hydrostatic Pressure Force on a Curved Surface
Pinching-off of Coated Vesicles
Bearing Stress
Due to the intricacy of these microforces, an average value, known as bearing stress, is often used by...