Purpose:
The hyperpolarized gases (3)He and (129)Xe have distinct properties and provide unique and complementary functional information from the lungs. A triple-nuclear, same-breath imaging examination of the lungs with (1)H, (3)He, and (129)Xe can therefore provide exclusive functional information from the gas images. In addition, the (1)H images provide complementary co-registered structural information in the same physiological time frame. The goal of this study was to design an RF system for triple nuclear lung MRI at 1.5T, consisting of a dual-tuned transceiver coil for (3)He and (129)Xe, RF switches and a nested (1)H receiver array.
Methods:
A dual-tuned transmit-receive dual-Helmholtz RF coil for (3)He and (129)Xe was designed and constructed to work in unison with a nested (1)H receiver array.
Results:
Triple-nuclear imaging (structural and ventilation) and apparent diffusion coefficient mapping of the human lungs was performed in the same breath-hold using the integrated RF system. B1 maps and volumetric ventilation imaging using a three-dimensional, balanced steady-state free precession pulse sequence performed with both hyperpolarized (3)He and (129)Xe indicate good stand-alone performance of the coil for the respective nucleus.
Conclusion:
Triple-nuclear same-breath lung imaging with a dual-tuned coil ((3)He and (129)Xe) and a nested (1)H array has been demonstrated with a custom RF system.
More Related Videos
09:08Author Spotlight: Standardization and Best Practices for Advancing Lung Imaging Using 129Xe MRI
Published on: November 21, 2023
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
Related Concept Videos
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Magnetic Resonance Imaging
¹H NMR: Complex Splitting
Splitting diagrams or splitting tree diagrams are routinely used to depict such complex couplings. While drawing splitting diagrams, the splitting with the larger coupling constant is usually applied...