Related Experiment Video
Updated: Oct 25, 2026

Use of a Multi-compartment Dynamic Single Enzyme Phantom for Studies of Hyperpolarized Magnetic Resonance Agents
Published on: April 15, 2016
HARP: HARmonizing In-Vivo Single-Shell Diffusion MRI Using Phantom-Only Training
Hwihun Jeong1, Qiang Liu1,2, Kathryn E Keenan3
1Department of Psychiatry, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA.
Purpose:
Combining multi-site diffusion MRI (dMRI) data is hindered by inter-scanner variability, which confounds subsequent analysis. Previous harmonization methods require large, matched or traveling human subjects from multiple sites, which are impractical to acquire in many situations. This study aims to develop a deep learning-based dMRI harmonization framework that eliminates the reliance on multi-site in-vivo traveling human data for training.
Methods:
HARP employs a voxel-wise 1D neural network trained on an easily transportable diffusion phantom. The model learns relationships between spherical harmonics coefficients of different sites without memorizing spatial structures.
Results:
HARP reduced inter-scanner variability levels significantly in various measures. Quantitatively, it decreased inter-scanner variability as measured by standard error in FA (12%), MD (10%), and GFA (30%) with scan-rescan standard error as the baseline, while preserving fiber orientations and tractography after harmonization.
Conclusion:
We believe that HARP represents an important first step toward dMRI harmonization using only phantom data, thereby obviating the need for complex, matched in vivo multi-site cohorts. This phantom-only strategy substantially enhances the feasibility and scalability of quantitative dMRI for large-scale clinical studies.