Calibration system for measuring the radon flux density
A Onishchenko1, M Zhukovsky2, V Bastrikov2
1Institute of Industrial Ecology UB RAS, Ekaterinburg, Russia onischenko@ecko.uran.ru.
Radiation Protection Dosimetry
|May 16, 2015
Summary
Accurate radon flux measurements are crucial for assessing radon-prone areas and monitoring uranium mining releases. Calibration of accumulation chambers and activated charcoal chambers ensures reliable radon flux data.
Area of Science:
- Environmental Science
- Geophysics
- Radiological Protection
Background:
- Radon flux measurement from soil is vital for assessing radon-prone areas and monitoring releases from uranium mining.
- Accumulation chambers and activated charcoal chambers are commonly used devices for radon flux measurements.
- Measurement accuracy is influenced by chamber geometry and radon diffusion in soil.
Purpose of the Study:
- To construct a calibration system for attesting radon flux measurement devices.
- To calibrate accumulation chambers and activated charcoal chambers.
- To validate measurement results using 2D modeling.
Main Methods:
- Construction of a calibration system for radon flux measurement devices.
- Conducting calibration measurements using accumulation chambers and activated charcoal chambers.
- Utilizing 2D modeling for comparison with measurement results.
Main Results:
- A calibration system for radon flux measurement devices was successfully established.
- Calibration measurements were performed on both types of chambers.
- Observed good agreement between 2D modeling predictions and experimental measurement results.
Conclusions:
- Reliable radon flux measurements can be achieved using activated charcoal chambers (approx. 75 L) or hollow headspace accumulation chambers (7-10 L with height >15 cm).
- The developed calibration system and 2D modeling aid in ensuring measurement accuracy.
- Optimized chamber design is key to obtaining dependable radon flux data.
Related Concept Videos
Flame Photometry: Lab
1.3K
In a flame photometer, when a solution like potassium chloride is aspirated into the flame, the solvent evaporates, leaving behind dehydrated salt. This salt dissociates into free gaseous atoms in their ground state. Some of these atoms absorb energy from the flame, leading to their excitation. The excited atoms return to the ground state, emitting photons at characteristic wavelengths. Because only electronic transitions are involved, the resulting emission lines are very narrow. The intensity...
1.3K
Constant Volume Calorimetry
31.8K
Calorimeters are useful to determine the heat released or absorbed by a chemical reaction. Coffee cup calorimeters are designed to operate at constant (atmospheric) pressure and are convenient to measure heat flow (or enthalpy change) accompanying processes that occur in solution at constant pressure. A different type of calorimeter that operates at constant volume, colloquially known as a bomb calorimeter, is used to measure the energy produced by reactions that yield large amounts of heat and...