Development and validation of a method using supported liquid extraction for aldosterone determination in human
Cécile Meunier1, Dominique Blondelle1, Patrice Faure2
1Unit of Hormone and Nutrition, Department of Biochemistry, Toxicology and Pharmacology, University Hospital, Grenoble, France.
Background:
Accurate quantitation of aldosterone is essential for screening, diagnosis and subtype classification in primary aldosteronism. A simple, sensitive method for aldosterone in human plasma using supported liquid extraction (SLE) in combination with liquid chromatography tandem mass spectrometry (LC-MS/MS) was developed and validated.
Methods:
Plasma samples were diluted with water containing d7-aldosterone as internal standard. The samples were extracted with methyl-tert-butyl-ether (MTBE) on SLE cartridges. Separation was carried out on a Luna C18 (2) column using a methanol-water gradient. Detection was performed in the negative electrospray multiple reaction monitoring (MRM) quantitation. The use of water-based calibrators was evaluated against calibrators prepared in steroid-free serum.
Results:
The assay was linear up to 3265pmol/L with an LOQ of approximately 40pmol/L. Within-run and between-run precision for plasma aldosterone were less than 10% except at low level near LOQ but were still less than 14.7% (Westgard's desirable specification). The mean recovery of the analyte added to plasma was greater than 97.7% and matrix effects were less than 4%. Comparison with another LC-MS/MS method was performed on a more sensitive instrument (ABSciex TQ 5500) and gave the equation API 3000=0.957×TQ 5500+12.6, linear regression r(2)=0.974 (n=43). An estimation of the reference interval for adults was established on a group of healthy volunteers (n=53). Calibration with water-based calibrators was validated and can be used for measurement of aldosterone by LC-MS/MS.
Conclusions:
This method is reliable, easy to perform on plasma specimens in a clinical environment and is attractive because of its simplicity.