Related Experiment Video
Updated: Apr 11, 2026

Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Effect of repeated sterilization cycles on the physical properties of scaling instruments: a scanning electron
Alessandra Nogueira Porto1, Álvaro Henrique Borges1, Alex Semenoff-Segundo1
1Professor, Department of Master Program in Integrated Dentistry Science, University of Cuiabá (UNIC), Cuiabá, MT, Brazil.
Background:
Repeated sterilizations cycles cause undesirable alterations in the material properties of the instruments, such as corrosion, alterations in the hardness of the metal and the loss of the cutting sharpness of the instrument. This research examined the effect of repeated dry heat sterilization and autoclaves cycles on carbon steel (CS) and stainless steel (SS) curettes during the scaling and root planning.
Materials And Methods:
A total of 77 Gracey curettes were used in this study. Of these, 35 were SS and 42 were CS curettes submitted in different process: Dry heat, autoclave, inhibition of corrosion and autoclave, scaling, root planning and dry heat, scaling, root planning, inhibition of corrosion and autoclave. The inhibition of corrosion used on the carbon curettes (prior to sterilization in the autoclave) was sodium nitrite at 2%. The curettes received 10 consecutive cycles of sterilization and after that the cutting edges were examined in the electronic microscope, at 60 and 100 magnification times.
Results:
The images were evaluated by three independent examiners, who compared the photographs of each group with the control group.
Conclusion:
The surface corrosion products and a deterioration of the edges were observed and the results showed that the SS curettes suffered little alteration with sterilization, scaling, root planning whereas the CS curettes were visibly affected by sterilization in the autoclave, but when the inhibition of corrosion was used prior to the sterilization, the oxidation was considerably reduced.
Related Concept Videos
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Preparation of Samples for Electron Microscopy
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...