Do breath tests really underestimate blood alcohol concentration?
Journal of Analytical Toxicology
|March 1, 1989
Summary
Breathalyzer tests may overestimate blood alcohol concentration (BAC) in the postabsorptive state. Reanalysis shows significant underestimations and overestimations, with peak BAC occurring later than previously thought.
Area of Science:
- Forensic toxicology
- Analytical chemistry
Background:
- Breathalyzer devices are commonly used to estimate blood alcohol concentration (BAC).
- Previous studies suggested underestimation of BAC by breath tests in the postabsorptive state.
Purpose of the Study:
- To re-evaluate the accuracy of breath test results for blood alcohol concentration (BAC) in the postabsorptive state.
- To investigate the timing of the postabsorptive state relative to alcohol consumption.
Main Methods:
- Reanalysis of existing data on breath test versus actual BAC measurements.
- Comparison of reanalyzed data with other relevant studies.
Main Results:
- Reanalysis indicated 77% of subjects had their BAC underestimated, and 23% overestimated, contradicting initial reports.
- Further analysis revealed 68% underestimated, 16% were acceptably close, and 16% overestimated.
- The fully postabsorptive state may occur later than three hours after drinking.
Conclusions:
- Breath test results may overestimate actual BAC for extended periods after peak levels are reached.
- The timing of the postabsorptive state significantly impacts breathalyzer accuracy.
- Rethinking the interpretation of breathalyzer results in the postabsorptive phase is warranted.