Accuracy and precision assessment of a new blood glucose monitoring system
Background:
Blood glucose self-monitoring by individuals with diabetes is essential in controlling blood glucose levels. The International Organization for Standardization (ISO) introduced new standards for blood glucose monitoring systems (BGMS) in 2013 (ISO 15197: 2013). The CONTOUR PLUS® (CONTOUR PLUS) BGMS has been demonstrated to meet the 2013 ISO standards; however, no Chinese data on CONTOUR PLUS accuracy and precision have been published.
Methods:
This study evaluated the accuracy and precision of CONTOUR PLUS BGMS in quantitative glucose testing of capillary and venous whole blood samples obtained from 363 patients at three different hospitals.
Results:
Results of fingertip and venous blood glucose measurements by the CONTOUR PLUS system were compared with laboratory reference values to determine accuracy. Accuracy was 98.1% (96.06%-99.22%) for fingertip blood tests and 98.1% (96.02%-99.21%) for venous blood tests. Precision was evaluated across a wide range of blood glucose values (5.1-17.2 mmol/L), testing three blood samples repeatedly 15 times with the CONTOUR PLUS blood glucose meter using test strips from three lots. All within-lot results met ISO criteria (i.e., SD<0.42 mmol/L for blood glucose concentration <5.55 mmol/L; CV<7.5% for blood glucose concentration ≥5.55 mmol/L). Between-lot variations were 1.5% for low blood glucose concentration, 2.4% for normal and 3.4% for high.
Conclusions:
Accuracy of both fingertip and venous blood glucose measurements by the CONTOUR PLUS system was >95%, confirming that the system meets ISO 15197: 2013 requirements.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Uncertainty in Measurement: Accuracy and Precision
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Accuracy and Precision