Whole blood hemolysis detection on the GEM® Premier™ 7000: performance across hemoglobin A, S, and F
Antoine Puravet1,2, Benjamin Bouthors1, Benjamin Rieu3
1Biochemistry and Molecular Genetics Department, Centre Hospitalier Universitaire Clermont-Ferrand, Clermont-Ferrand, France.
Objectives:
Point-of-care blood gas analyzers are essential in critical care settings but remain vulnerable to pre-analytical interferences, particularly hemolysis, which may lead to erroneous potassium results. The GEM® Premier™ 7000 incorporates an onboard hemolysis detection system based on multi-wavelength co-oximetry, allowing direct estimation of free hemoglobin in whole blood samples. This study evaluated the analytical performance of this system across different hemoglobin variants, including fetal hemoglobin (HbF) and hemoglobin S (HbS).
Methods:
Fifty patient samples undergoing routine arterial blood gas and biochemical testing were analyzed using the GEM® Premier™ 7000, and Atellica Solution analyzers. Experimental hemolysis was induced to evaluate the performance of the onboard hemolysis detection system. In a second step, hemoglobin overload experiments using HbA0, HbF, and HbS were performed in samples from 10 healthy volunteers to assess the robustness of hemolysis detection across different hemoglobin variants.
Results:
Experimental hemolysis significantly increased the hemolysis index and potassium concentrations measured on the GEM® Premier™ 7000, with median hemolysis index values rising from 1 to 6 and median potassium concentrations from 3.6 to 9.8 mmol/L (p<0.001). The GEM® Premier™ 7000 accurately detected increasing concentrations of free HbA0, HbS, and HbF throughout the overload experiments.
Conclusions:
The integrated hemolysis detection system reliably detected clinically significant hemolysis in whole blood samples, supporting more reliable point-of-care blood gas testing.