Selective reporting in trials of high risk cardiovascular devices: cross sectional comparison between premarket
Lee Chang1, Sanket S Dhruva2, Janet Chu3
1Department of Medicine, Massachusetts General Hospital, 55 Fruit Street, Boston, MA 02114, USA.
Objective:
To investigate characteristics of clinical trials and results on safety and effectiveness reported in US Food and Drug Administration (FDA) documents for recently approved high risk cardiovascular devices compared with the characteristics and results reported in peer reviewed publications.
Design:
A search of the publicly available FDA database was performed for all cardiovascular devices that received premarket approval from 1 January 2000 to 31 December 2010. For each study listed in the premarket approval documents, a Medline search was conducted to obtain the corresponding publication.
Main Outcome Measures:
Clinical trial characteristics, primary endpoints, and safety and efficacy results in the FDA documents and corresponding publications.
Results:
106 cardiovascular devices received premarket approval from 1 January 2000 to 31 December 2010. FDA premarket approval documents for these devices contained 177 studies, of which 86 (49%) had been published by 1 January 2013. These 86 publications corresponded to 60 distinct devices. The mean time from FDA approval to publication in a peer reviewed journal was 6.5 months (range -4.8-7.5 years). In 22 (26%) of the 86 compared studies the number of participants enrolled in the study differed in the FDA summary and the corresponding publications. Of 152 primary endpoints identified in the FDA documents, in the corresponding publications three (2%) were labeled as secondary, 43 (28%) were unlabeled, and 15 (10%) were not found. Among the primary results, 69 (45%) were identical, 35 (23%) were similar, 17 (11%) were substantially different, and 31 (20%) could not be compared.
Conclusions:
Many clinical trials for high risk cardiovascular devices approved by the FDA remain unpublished. Even when trials are published, the study population, primary endpoints, and results can differ substantially from data submitted to the FDA.
Related Concept Videos
Clinical Trials: Overview
Clinical Trials
There are four phases in a clinical trial. A phase one...
Pharmacovigilance
This process, termed pharmacovigilance, aims to detect, evaluate, and minimize harmful effects related to medication use. The data collection for pharmacovigilance depends on spontaneous reporting systems, where healthcare professionals or patients voluntarily report suspected ADRs.
In some cases, there...
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Acute Coronary Syndrome III: Diagnostic Studies
Cardiovascular Drugs: Classification based on Therapeutic Indications