CLSI-based transference of the CALIPER database of pediatric reference intervals to Beckman Coulter DxC biochemical

Petra A T Araújo1, Dylan Thomas1, Tara Sadeghieh1

  • 1CALIPER Program, Paediatric Laboratory Medicine, The Hospital for Sick Children, Toronto, ON, Canada.

Clinical Biochemistry
|June 14, 2015
PubMed

Insights

Pediatric reference intervals from the CALIPER database were successfully transferred to Beckman Coulter DxC800 assays for most biochemical markers. This allows broader use of these vital pediatric reference intervals in clinical settings.

Area of Science:

  • Clinical Chemistry
  • Pediatric Laboratory Medicine
  • Biochemical Assays

Background:

  • The CALIPER program provides extensive pediatric reference intervals for biochemical markers, primarily using Abbott ARCHITECT assays.
  • Broader clinical application necessitates evaluating assay transference to other platforms, such as Beckman Coulter.

Purpose of the Study:

  • To assess the transferability of CALIPER pediatric reference intervals to 36 Beckman Coulter Synchron Unicel DxC800 assays.
  • To validate the transferred reference intervals using established clinical laboratory guidelines (CLSI C28-A3/EP9-A3).

Main Methods:

  • Compared 36 biochemical assays using 200 pediatric serum specimens on Abbott ARCHITECT ci8200 and Beckman Coulter DxC800 platforms.
  • Evaluated assay correlation using R(2) values, regression analysis, standardized residuals, Bland-Altman, and quantile-quantile plots.
  • Verified transferred reference intervals by analyzing samples from 100 healthy children on the Beckman Coulter system.

Main Results:

  • Reference intervals for 31 out of 36 biochemical assays were transferable to Beckman Coulter DxC800 assays.
  • Newly calculated intervals were verified for 28 out of 31 transferable assays.
  • Eighteen assays showed excellent correlation (R(2)≥0.95), and 13 showed strong correlation (0.77≤R2≤0.94).

Conclusions:

  • Successful transference of numerous biochemical markers from the CALIPER database to Beckman Coulter DxC800 assays was achieved.
  • This transference facilitates wider utilization of CALIPER pediatric reference intervals at institutions using Beckman Coulter DxC assays.