High-fidelity digital recording and playback sphygmomanometry system: device description and proof of concept
Jongshill Lee1, Youngjoon Chee, Inyoung Kim
1aDepartment of Biomedical Engineering, Hanyang University, Seoul bDepartment of Biomedical Engineering, University of Ulsan, Ulsan, Korea cHypertension Center, STRIDE-7, Third University Department of Medicine, Sotiria Hospital, Athens, Greece dMedaval Ltd eConway Institute of Biomolecular and Biomedical Research, University College Dublin, Dublin, Ireland.
Objectives:
This study describes the development of a new digital sphygmocorder (DS-II), which allows the digital recording and playback of the Korotkoff sounds, together with cuff pressure waveform, and its performance in a pilot validation study.
Materials And Methods:
A condenser microphone and stethoscope head detect Korotkoff sounds and an electronic chip, dedicated to audio-signal processing, is used to record high-quality sounds. Systolic and diastolic blood pressure (SBP/DBP) are determined from the recorded signals with an automatic beat detection algorithm that displays the cuff pressure at each beat on the monitor. Recordings of Korotkoff sounds, with the cuff pressure waveforms, and the simultaneous on-site assessments of SBP/DBP were performed during 100 measurements in 10 individuals. The observers reassessed the recorded signals to verify their accuracy and differences were calculated.
Results:
The features of the high-fidelity DS-II, the technical specifications and the assessment procedures utilizing the playback software are described. Interobserver absolute differences (mean±SD) in measurements were 0.7±1.1/1.3±1.3 mmHg (SBP/DBP) with a mercury sphygmomanometer and 0.3±0.9/0.8±1.2 mmHg with the DS-II. The absolute DS-II mercury sphygmomanometer differences were 1.3±1.9/1.5±1.3 mmHg (SBP/DBP).
Conclusion:
The high-fidelity DS-II device presents satisfactory agreement with simultaneous measurements of blood pressure with a mercury sphygmomanometer. The device will be a valuable methodology for validating new blood pressure measurement technologies and devices.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure