A Highly Specific, Fully Validated Urinary Ethyl Glucuronide Analysis Using Solid Phase Extraction-Liquid
Munevver Acikkol1, Ali Acar Ozdemir, Beril Anilanmert
1Istanbul University, Institute of Forensic Sciences, 34303, Cerrahpasa, Istanbul, Turkey.
A new, validated LC/MS/MS method accurately quantifies ethyl glucuronide (EtG) in urine, serving as a reliable biomarker for alcohol consumption monitoring in treatment programs and forensic toxicology.
Area of Science:
- Analytical Chemistry
- Forensic Toxicology
- Clinical Chemistry
Background:
- Ethyl glucuronide (EtG) is a direct metabolite of ethanol, detectable in urine days after alcohol consumption.
- Accurate EtG quantification is crucial for monitoring alcohol abstinence in treatment programs and in forensic investigations.
- Existing methods may require improvement in specificity and sensitivity for reliable EtG detection.
Purpose of the Study:
- To develop and validate a highly specific and selective liquid chromatography-tandem mass spectrometry (LC/MS/MS) method for quantitative EtG determination in urine.
- To establish a reliable analytical tool for monitoring alcohol use, particularly in clinical and forensic settings.
Main Methods:
- Development and validation of an LC/MS/MS method for ethyl glucuronide (EtG) quantification in urine.
- Utilized propyl glucuronide (PrG) as an internal standard to enhance method selectivity.
- Employed solid-phase extraction (SPE) for sample cleanup and a C18 column for chromatographic separation.
- Validated parameters included selectivity, linearity, LOD, LOQ, precision, and recovery.
Main Results:
- The developed LC/MS/MS method demonstrated high specificity and selectivity for EtG determination.
- The limit of detection (LOD) was 90.8 ng/mL and the limit of quantification (LOQ) was 185.0 ng/mL.
- Intraday and interday precision were within acceptable limits, confirmed by HorRat values.
- The method showed excellent performance in real-world sample analysis.
Conclusions:
- The validated LC/MS/MS method provides a reliable and accurate solution for quantifying EtG in urine.
- This method is suitable for monitoring patients undergoing alcohol addiction treatment.
- The assay is applicable to both antemortem and postmortem toxicological investigations, aiding in alcohol use assessment.
More Related Videos
14:48Proteomic Profile of EPS-Urine through FASP Digestion and Data-Independent Analysis
Published on: May 8, 2021
09:49Use of a Battery of Chemical and Ecotoxicological Methods for the Assessment of the Efficacy of Wastewater Treatment Processes to Remove Estrogenic Potency
Published on: September 11, 2016
Related Concept Videos
Ion-Exchange Chromatography
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...