Related Experiment Video
Updated: Apr 9, 2026

Multiplexed Fluorescent Microarray for Human Salivary Protein Analysis Using Polymer Microspheres and Fiber-optic Bundles
Published on: October 10, 2013
A rapid sandwich immunoassay for human fetuin A using agarose-3-aminopropyltriethoxysilane modified microtiter plate
Sandeep Kumar Vashist1, E Marion Schneider2, John H T Luong3
1Laboratory for MEMS Applications, Department of Microsystems Engineering-IMTEK, University of Freiburg, Georges-Koehler-Allee 103, 79110 Freiburg, Germany; HSG-IMIT-Institut für Mikro- und Informationstechnik, Georges-Koehler-Allee 103, 79110 Freiburg, Germany.
Abstract:
A rapid sandwich immunoassay (IA) with enhanced signal response for human fetuin A (HFA) was developed by modifying the surface of a KOH-treated polystyrene microtiter plate (MTP) with agarose and 3-aminopropyltriethoxysilane (APTES). The agarose-APTES complex binds covalently to the hydroxyl moiety of the MTP plate to serve as a binding platform for bioconjugation of EDC-activated anti-HFA antibody (Ab) via carbodiimide coupling. The one-step kinetics-based sandwich enzyme-linked immunosorbent assay (ELISA) enabled the detection of HFA in 30 min with a limit of detection (LOD) and a linear range of 0.02 ng mL(-1) and 1-243 ng mL(-1), respectively. It detected HFA spiked in diluted human whole blood and serum, and HFA in ethylenediaminetetraacetic acid (EDTA)-plasma of patients with high precision similar to that of conventional ELISA. The anti-HFA Ab-bound agarose-functionalized MTPs retained their functional activity after 6 weeks of storage in 0.1 M PBS, pH 7.4 at 4 °C.