Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Assessment of Diffusion and Perfusion01:17

Assessment of Diffusion and Perfusion

2.1K
Understanding and evaluating diffusion and perfusion is critical in assessing a patient's respiratory and circulatory health. These processes play key roles in maintaining the body's internal environment, ensuring that tissues receive adequate oxygen while waste products are efficiently removed.
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
2.1K
Magnetic Resonance Imaging01:24

Magnetic Resonance Imaging

10.4K
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
10.4K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Tumor-associated epilepsy at diagnosis in glioblastoma patients reveals an inflammatory molecular signature and is associated with better overall survival.

Neuro-oncology·2026
Same author

Multispectral 7 Tesla MRI as a potential predictor of dopamine transporter deficiency in Parkinson's disease.

Imaging neuroscience (Cambridge, Mass.)·2026
Same author

Imaging patterns of glioblastoma progression: Comparative analysis of sequential proton boost versus conventional photon therapy.

Neuro-oncology advances·2026
Same author

Brain morphology in Anorexia Nervosa and its subtypes: A multi-cohort study of individual participant data.

PLoS medicine·2026
Same author

OpenMRF: A Modular, Vendor-Neutral Open-Source Framework for Reproducible Magnetic Resonance Fingerprinting using Pulseq.

ArXiv·2026
Same author

Mitochondrial respirasome-like supercomplexes support metabolic flexibility in yeast.

Nature communications·2026
Same journal

Influence of gadolinium-based contrast agent (GBCA) on the diffusion weightings of breast lesions: an intra-patient analysis.

Magma (New York, N.Y.)·2026
Same journal

Evaluation of the diffusion time dependence of the IVIM effect based on realistic capillary flow simulations in mouse brain.

Magma (New York, N.Y.)·2026
Same journal

An evaluation of brain volume and cortical thickness measurement at 0.55 T.

Magma (New York, N.Y.)·2026
Same journal

Net zero emission MR imaging using a permanent 0.4 T magnet.

Magma (New York, N.Y.)·2026
Same journal

Special issue on "deuterium metabolic imaging".

Magma (New York, N.Y.)·2026
Same journal

Black-blood dynamic contrast-enhanced MRI of abdominal aortic aneurysms.

Magma (New York, N.Y.)·2026
See all related articles

Related Experiment Video

Updated: Apr 9, 2026

Diffusion Imaging in the Rat Cervical Spinal Cord
10:46

Diffusion Imaging in the Rat Cervical Spinal Cord

Published on: April 7, 2015

12.3K

Acoustic-noise-optimized diffusion-weighted imaging.

Martin Ott1, Martin Blaimer2, David M Grodzki3

  • 1Research Center for Magnetic-Resonance-Bavaria, Am Hubland, 97074, Würzburg, Germany. mott@physik.uni-wuerzburg.de.

Magma (New York, N.Y.)
|June 21, 2015
PubMed
Summary
This summary is machine-generated.

This study introduces a new method to lower the loud, disruptive noise produced during brain and body scans using diffusion-weighted MRI. By adjusting how the machine's internal magnets move, researchers successfully reduced noise levels by 20 decibels, making the experience much quieter for patients while keeping image quality high.

Keywords:
Acoustic noise reductionDWIPatient comfortQuiet MRIReadout-segmented EPIMRI safetygradient switchingauditory comfortecho-planar imaging

Frequently Asked Questions

More Related Videos

Advanced Diffusion Imaging in The Hippocampus of Rats with Mild Traumatic Brain Injury
10:33

Advanced Diffusion Imaging in The Hippocampus of Rats with Mild Traumatic Brain Injury

Published on: August 14, 2019

9.1K
Co-analysis of Brain Structure and Function using fMRI and Diffusion-weighted Imaging
17:06

Co-analysis of Brain Structure and Function using fMRI and Diffusion-weighted Imaging

Published on: November 8, 2012

27.2K

Related Experiment Videos

Last Updated: Apr 9, 2026

Diffusion Imaging in the Rat Cervical Spinal Cord
10:46

Diffusion Imaging in the Rat Cervical Spinal Cord

Published on: April 7, 2015

12.3K
Advanced Diffusion Imaging in The Hippocampus of Rats with Mild Traumatic Brain Injury
10:33

Advanced Diffusion Imaging in The Hippocampus of Rats with Mild Traumatic Brain Injury

Published on: August 14, 2019

9.1K
Co-analysis of Brain Structure and Function using fMRI and Diffusion-weighted Imaging
17:06

Co-analysis of Brain Structure and Function using fMRI and Diffusion-weighted Imaging

Published on: November 8, 2012

27.2K

Area of Science:

  • Medical imaging physics within diagnostic radiology
  • Acoustic-noise-optimized diffusion-weighted imaging research in biomedical engineering

Background:

High sound pressure levels during magnetic resonance imaging scans remain a significant challenge for clinical environments. Standard protocols often exceed 100 decibels, creating discomfort for individuals undergoing diagnostic procedures. No prior work had resolved the trade-off between rapid data acquisition and auditory safety. That uncertainty drove the need for quieter scanning sequences. Prior research has shown that gradient switching is the primary source of these intense sounds. This gap motivated the development of alternative pulse sequences that prioritize auditory comfort. Existing literature highlights the difficulty of maintaining diagnostic clarity while slowing down hardware movements. The current study addresses these limitations by refining sequence parameters to minimize sound production.

Purpose Of The Study:

This work aimed to decrease the intense acoustic noise generated during diffusion-weighted magnetic resonance imaging. Clinical practice often involves sound levels exceeding 100 decibels, which poses challenges for patient well-being. The researchers sought to develop a sequence that maintains diagnostic quality while significantly lowering auditory output. This gap motivated the exploration of alternative gradient control strategies. The team focused on replacing rapid k-space coverage with slower, optimized hardware movements. That uncertainty drove the need for a systematic evaluation of noise-reduction techniques. The study specifically targets the mechanical vibrations caused by high-speed gradient switching. By refining these parameters, the authors intended to create a more comfortable environment for individuals undergoing diagnostic scans.

Main Methods:

The investigators designed a modified readout-segmented echo-planar imaging sequence to test noise reduction. Review approach involved comparing this new protocol against standard single-shot imaging techniques. Hardware adjustments focused on utilizing smaller segment widths to limit gradient activity. All auxiliary magnetic field pulses were also tuned for slower movement profiles. Imaging experiments were performed on both human volunteers and clinical patients. Sound levels were recorded and evaluated across four distinct measurement configurations. Testing occurred at both 1.5 Tesla and 3 Tesla field strengths to ensure broad applicability. Data analysis confirmed the feasibility of the proposed sequence for routine diagnostic tasks.

Main Results:

The researchers achieved a maximum sound reduction of 20 decibels compared to standard protocols. This decrease represents a fourfold improvement in auditory perception for the patient. Image quality remained consistent with traditional single-shot echo-planar imaging sequences. Key findings from the literature indicate that this improvement requires a trade-off in efficiency. Scan durations increased by 27 to 54 percent relative to conventional methods. The results demonstrate that quieter scanning is possible without compromising diagnostic accuracy. These measurements were consistent across both 1.5T and 3T field strengths. The data confirm that hardware optimization effectively lowers the intense vibrations inherent in magnetic resonance imaging.

Conclusions:

The authors propose that their modified sequence effectively lowers sound pressure during magnetic resonance procedures. This approach achieves a twenty decibel decrease, which translates to a fourfold reduction in perceived loudness. Synthesis and implications suggest that patient comfort is improved without sacrificing diagnostic utility. The researchers maintain that image quality remains comparable to traditional single-shot methods. They acknowledge that scan times increase by twenty-seven to fifty-four percent using this technique. Future efforts should involve larger cohorts to fully evaluate subjective improvements in the clinical experience. The team concludes that this method represents a viable alternative for noise-sensitive individuals. These findings provide a clear path for integrating quieter protocols into routine practice.

The researchers utilized a readout-segmented echo-planar imaging sequence with reduced segment widths. This configuration allows for lower gradient slew rates and amplitudes, which directly minimizes the mechanical vibrations responsible for loud sounds during the scanning process.

The team employed a readout-segmented echo-planar imaging sequence. This specific tool enables the manipulation of k-space coverage, allowing for the substitution of rapid gradient switching with slower, more controlled hardware movements to achieve quieter operation.

Lower slew rates are necessary because they reduce the rapid acceleration of magnetic field gradients. This technical adjustment prevents the intense mechanical stress that typically generates high-decibel acoustic output in standard imaging hardware.

The researchers analyzed four distinct diffusion-weighted imaging protocols at both 1.5T and 3T field strengths. These data types allowed the team to quantify the acoustic benefits across different clinical hardware configurations.

The study measured a reduction of up to 20 dB(A) in acoustic output. This specific phenomenon corresponds to a fourfold decrease in how humans perceive the loudness of the magnetic resonance scanner.

The authors propose that this method will increase patient comfort during examinations. They suggest that while the technique is feasible, larger trials are needed to fully characterize the subjective benefits for those undergoing these scans.