This study introduces a new laparoscopic ultrasound (LUS) image guidance system for liver surgery. The system uses locally rigid registration to accurately guide surgeons, showing promising results in phantom and animal studies for improved surgical navigation.
Area of Science:
- Medical Imaging
- Surgical Technology
- Computational Anatomy
Background:
- Laparoscopic liver resection offers benefits but faces challenges due to limited visualization and lack of tactile feedback.
- Image guidance is a potential solution, but organ deformation in soft tissues like the liver complicates registration.
- Accurate image guidance is crucial for overcoming the limitations of laparoscopic liver surgery.
Purpose of the Study:
- To propose and evaluate a laparoscopic ultrasound (LUS) image guidance system for liver surgery.
- To investigate the feasibility of locally rigid registration for aligning preoperative models with intraoperative ultrasound data.
- To assess the system's accuracy in a phantom and in vivo porcine models.
Main Methods:
- Developed a real-time segmentation method for extracting vessel center points from 2D LUS images.
- Utilized electromagnetically tracked LUS and preoperative CT-derived vessel centerline models.
- Employed landmark-based and iterative closest point (ICP) point-to-line registration techniques for aligning models.
Main Results:
- The locally rigid ICP method achieved an RMS residual error of 0.7 mm in phantom studies.
- Mean target registration error (TRE) in porcine studies was 3.58 mm and 2.99 mm.
- Landmark-based registration yielded mean TREs of 4.23 mm and 6.57 mm, depending on pneumoperitoneum conditions.
Conclusions:
- A practical image-guided surgery system using locally rigid registration of CT models to LUS-identified vasculature was developed.
- The system demonstrated accuracy suitable for surgical requirements in phantom and porcine liver resection models.
- Locally rigid registration is deemed sufficient for practical image guidance in laparoscopic liver surgery in the near future.