Background:
Barium esophagograms have poor sensitivity in detecting leaks. We hypothesized that heating barium would decrease viscosity, facilitate extravasation, and enhance its sensitivity in detecting esophageal leaks.
Methods:
We characterized the viscosity of barium at increasing temperatures. We measured the radiopacity of barium at 25°C and 50°C. We determined the smallest diameter defect in esophagus that barium can detect by perforating a porcine esophageal segment with angiocatheters of various diameters, injecting barium at 25°C, and observing extravasation of contrast. We repeated this with barium heated to 30°C, 40°C, 50°C, and 70°C. To determine the ability of barium to detect a staple line leak, we perforated a stapled esophageal segment by air insufflation, injected barium at different temperatures, and monitored extravasation. We used Visipaque, a water-soluble contrast agent, for comparison in all experiments.
Results:
The viscosity of barium decreased with increasing temperature. The radiopacity of barium did not change with increasing temperature and was higher than that of Visipaque (P < 0.001). The size of the smallest detectable leak decreased from 2.1 mm with barium at 25°C to 1.3 mm at 40°C and 1.1 mm with Visipaque (P < 0.0001). The sensitivity of staple line leak detection increased from 0% for barium at 25°C to 80% (P = 0.02) with barium at 40°C. There was no significant difference in sensitivity between barium at 40°C and Visipaque.
Conclusions:
Barium warmed to 40°C offers the best sensitivity of esophageal leak detection without compromising radiopacity. Barium at 40°C may be the optimum choice for swallow study to detect esophageal leaks.
More Related Videos
Related Concept Videos
Upper GI Series: Barium Swallow
Purpose and Procedure
Patients undergoing this procedure ingest a liquid containing barium sulfate with a chalky...
Lower GI Series: Barium Enema
Procedure Details
The examination begins by inserting a lubricated rectal tube into the patient's rectum to administer a radiopaque barium solution. The barium flow is carefully...
Barrett Esophagus-II: Clinical Manifestations and Management
To diagnose Barrett's esophagus, healthcare providers often recommend an endoscopy for those showing symptoms of acid reflux. The procedure...