Prevalence of Sensor Saturation in Wheelchair Seat Interface Pressure Mapping
Michael Wininger1, Barbara A Crane
1a University of Hartford , Department of Rehabilitation Sciences , West Hartford , Connecticut.
Abstract:
Pressure mapping is a frequently used tool with great power to provide information about the forces between a patient and a wheelchair seat. One widely recognized limitation to this paradigm is the possibility of data loss due to sensor saturation. In this study, we seek to quantify and describe the saturation observed in the measurement of interface pressures of wheelchair users. We recorded approximately two minutes of interface pressure data from 22 elderly wheelchair users (11M/11F, 80 ± 10 years) and found that 4.7% of data frames had 1 saturated sensor, and 9.0% had more than one saturated sensor, for a total of 13.7% of all frames of data. Data from three of the 22 subjects (13.6%) were substantially affected by the persistent presence of saturated sensors. We conclude that for this population of elderly wheelchair users, sensor saturation may be a concern and should be factored properly into study design a priori.