Waqas Ullah1, Ross J Hunter2, Victoria Baker2
1From the Cardiovascular Biomedical Research Unit, St. Bartholomew's Hospital, Barts Health NHS Trust, London, United Kingdom (W.U., R.J.H., V.B., M.B.D., S.S., M.J.E., R.J.S.); and Department of Cardiovascular Medicine, Alfred Hospital and Baker IDI Heart and Diabetes Institute, Melbourne, Victoria, Australia (L.-H.L.). richard.schilling@bartshealth.nhs.uk.
Insights
Optimal catheter contact force (CF) during left atrial mapping is crucial for accurate electrograms. Maintaining CF at or above 10 g minimizes variations and ensures reliable complex fractionated atrial electrogram data for ablation procedures.
Area of Science:
- Electrophysiology
- Cardiology
- Medical Device Technology
Background:
- Optimal catheter contact parameters for left atrial mapping are not well-established.
- Variations in catheter contact can affect electrogram data quality.
- Understanding these parameters is vital for successful atrial fibrillation ablation.
Purpose of the Study:
- To determine the impact of catheter contact force (CF) and orientation on left atrial electrogram parameters.
- To establish optimal CF thresholds for minimizing data variability during mapping.
- To guide best practices for catheter manipulation during electrophysiological procedures.
Main Methods:
- Analysis of 1965 stable mapping points from 30 patients undergoing atrial fibrillation ablation.
- Collected bipolar electrogram and 8-second CF data during both atrial fibrillation and sinus rhythm.
- Examined electrogram characteristics at varying CF levels and catheter orientations.
Main Results:
- Higher CF generally led to less positive electrogram complexes.
- Increasing CF significantly increased complex size when initial CF was below 10 g.
- Catheter orientation changes impacted complex morphology but not spectral parameters.
Conclusions:
- Catheter contact force and orientation significantly influence left atrial electrogram mapping results.
- A minimum CF of 10 g is recommended to mitigate the influence of CF variations.
- These findings aid in standardizing mapping techniques for improved ablation outcomes.
Background:
During left atrial mapping, optimal contact parameters minimizing variation secondary to catheter contact are not established.
Methods And Results:
Across 30 patients undergoing first-time atrial fibrillation ablation, 1965 stable mapping points (1409 atrial fibrillation, 556 sinus rhythm), comprising 8-s contact force (CF) and bipolar electrogram data were analyzed. Points were taken in groups at locations with CF or catheter orientation actively changed between acquisitions. Complexes were less positive at higher CF (Spearman ρ, -0.2; P<0.005, both rhythms). Increasing CF at a location significantly increased complex size, but only where initial CF was <10 g, and if the change was ≥4.5 g in sinus rhythm and ≥8 g in atrial fibrillation (P<0.0005, both rhythms): if initial CF was ≥10 g, no change was observed, regardless of CF change (P>0.05, both). Atrial ectopics during sinus rhythm were observed more frequently when CF was ≥10 g (P<0.0005). Increasing CF at a location was associated with an increase in the complex fractionated atrial electrogram interval confidence level score, but only if initial CF was <10 g and CF increased ≥8 g (P=0.003). The dominant frequency and organization index were unaffected by CF (P>0.1 for both). Changing catheter orientation from perpendicular to parallel in atrial fibrillation was associated with smaller, more positive complexes (P=0.001 for both), but no changes in complex fractionated atrial electrogram scores, dominant frequency or organization index (P>0.08 for each).
Conclusions:
During left atrial electrogram mapping, including complex fractionated atrial electrogram but not spectral parameter mapping, CF and catheter orientation influence results: consequently, mapping CFs should be ≥10 g to negate the influence of CF.
Clinical Trials Registration:
URL: http://clinicaltrials.gov/. Unique identifier: NCT01587404.
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Electrophysiology of Normal Cardiac Rhythm
ECG Interpretation of Arrhythmias I: Sinus Arrhythmias
Types of Arrhythmias
Sinus Node Arrhythmias
Sinus Bradycardia: Originating from the sinoatrial (SA) node, sinus bradycardia involves slower impulses, resulting in a heart rate of less than 60 beats per minute (bpm). Causes include sleep, vagal stimulation, beta-blockers, hypothyroidism,...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...