Accuracy of community-based blood pressure devices versus validated self-use devices
Objectives:
To examine the accuracy of a community-based blood pressure monitor (CBPM) versus a validated automatic blood pressure monitor (ABPM); to assess providers' recommendations regarding use of CBPMs; and to assess if potentially inaccurate blood pressure readings may result in treatment changes.
Methods:
A convenience sample of 50 participants was recruited and met criteria for a randomized, controlled, crossover design trial in a community pharmacy in January and February 2013. Participants completed a screening survey and were educated about how a pharmacist can assist in achieving blood pressure control. Blood pressure measurements were performed using a CBPM and a validated ABPM. Participants were asked to identify their primary care provider (PCP); if identified, the PCP was faxed a survey to complete. The difference in blood pressure readings between the two monitors was examined and PCP survey responses were examined.
Results:
There were significant differences between the means obtained from the two different monitors for systolic blood pressure (P = 0.006) and diastolic blood pressure (P = 0.012) measurements. Due to differences between readings, recommendations for change in treatment (or initiation of blood pressure medication) may have been different for 38% of participants. PCPs indicated that they often recommend CBPMs for self-monitoring.
Conclusion:
Patients, pharmacists, and providers should be aware of potential inaccuracies when using a CBPM for self-monitoring.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)