Accuracy of inpatient oxygen administration
A A Jeffrey1, S Ray, N J Douglas
1Department of Respiratory Medicine, City Hospital, Edinburgh.
Thorax
|December 1, 1989
Abstract:
The accuracy of supplemental oxygen delivery was assessed in two medical units. Only 90 of 206 patients (44%) were receiving oxygen as prescribed. The ward oxygen rotameters were inaccurate at normal clinical flow rates, the range of flows being delivered as a percentage of the indicated flow varying from 15% at 8 l/min to 40% at 1 and 2 l/min.