Simultaneous ex vivo Functional Testing of Two Retinas by in vivo Electroretinogram System
Published on: May 6, 2015
The effect of gold foil electrode position on the electroretinogram in human subjects
A McAllan1, J Sinn, G W Aylward
1Department of Ophthalmology, Sydney Eye Hospital, Woolloomooloo, NSW, Australia.
Vision Research
|January 1, 1989
Abstract:
Photopic ERGs were recorded in two human subjects using gold foil electrodes mounted at various positions in the lower fornix. The peak to peak amplitude of the b-wave was reduced in both subjects when the centre of the electrode was placed beyond 15 mm from the medial side of the limbus. The implicit time of the a-wave and the b-wave was not dependent on electrode position. These results emphasize the importance of correct placement of gold foil electrodes.