Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow
Published on: March 5, 2020
In Vitro Hemocompatibility Testing of Dyneema Purity Fibers in Blood Contact
Amir Basir1, Philip de Groot, Paul F Gründeman
1From the Departments of *Experimental Cardiology, †Clinical Chemistry and Haematology, ‡Vascular Surgery, and §Cardio-Thoracic Surgery, UMC Utrecht, Netherlands.
Objective:
Heart valve and vascular prosthesis implantation is a common procedure for patients with heart valve stenosis or regurgitation and dilated or obstructive vascular disease. Drawbacks of conventional valve prostheses are the requirement for anticoagulant drugs, moderate durability, and suboptimal resistance to fatigue and tear. Dyneema Purity fibers are made from ultra-high-molecular-weight polyethylene filaments and are very thin, flexible, and fatigue and abrasion resistant and have high strength. Therefore, prostheses made from Dyneema Purity fibers might be attractive for use in the minimally invasive treatment of valvular- and vascular diseases. The aim of this study was to test the hemocompatibility of Dyneema Purity fibers in contact with blood.
Methods:
Real-time platelet adhesion in human blood of 3 volunteers was quantified after 5 minutes of perfusion on single filaments (Ø 15 μm) of Dyneema Purity and polyester fibers. Plasma thrombin generation was measured by fluoroscopy for patches of Dyneema Purity fibers and for 5 commonly used polyester and expanded polytetrafluoroethylene cardiovascular prostheses.
Results:
Platelet adhesion per 1 mm was 6 ± 1.4 on Dyneema Purity filaments and 15 ± 3.4 on polyester filaments (P = 0.02). Total formed thrombin and the time to peak of its maximum were noninferior for patches of Dyneema Purity fibers compared with the reference materials.
Conclusions:
Dyneema Purity fibers are noninferior in adhesion and coagulation activation compared with commonly used cardiovascular prostheses.