Subject- and resource-specific monitoring and proactive management of parallel radiofrequency transmission
Related Experiment Video
Updated: Apr 6, 2026

Effective Analysis of Human Exposure Conditions with Body-worn Dosimeters in the 2.4 GHz Band
Published on: May 2, 2018
Cem M Deniz1,2,3,4, Leeor Alon1,2,3,4, Ryan Brown1,3
1Center for Advanced Imaging Innovation and Research (CAI2R) and Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University School of Medicine, New York, New York, USA.
Purpose:
Develop a practical comprehensive package for proactive management of parallel radiofrequency (RF) transmission.
Methods:
With a constrained optimization framework and predictive models from a prescan based multichannel calibration, we presented a method supporting design and optimization of parallel RF excitation pulses that accurately obey the forward/reflected peak and average power limits of the RF power amplifiers in parallel transmit imaging experiments and Bloch simulations. Moreover, local SAR limits were incorporated into the parallel RF excitation pulses using electromagnetic field simulations. Virtual transmit coils concept for minimization of reflected power (effecting subject-specific matching) was additionally demonstrated by leveraging experimentally calibrated power models.
Results:
Incorporation of experimentally calibrated power prediction models resulted in accurate compliance with prescribed hardware and global specific absorption rate (SAR) limits. Incorporation of spatial average 10 g SAR models, facilitated by simplifying numerical approximations, provided assurance of patient safety. RF pulses designed with various constraints demonstrated excellent excitation fidelity-the normalized root-mean-square error of the simulated excitation profiles was 2.6% for the fully constrained pulses, comparable to that of the unconstrained pulses. An RF shimming example showed a reduction of the reflected-to-forward power ratio to 1.7% from a conventional approach's 8.1%.
Conclusion:
Using the presented RF pulse design method, effective proactive management of the multifaceted power and SAR limits was demonstrated in experimental and simulation studies. Magn Reson Med 76:20-31, 2016. © 2015 Wiley Periodicals, Inc.