Objective assessment of robotic surgical skill using instrument contact vibrations
Ernest D Gomez1,2, Rajesh Aggarwal3,4, William McMahan5
1Department of Otorhinolaryngology - Head and Neck Surgery, Hospital of the University of Pennsylvania, Philadelphia, PA, USA. ernest.gomez@uphs.upenn.edu.
Background:
Surgical skill evaluation ordinarily requires tedious video review and survey completion, while new automatic approaches focus on evaluating the quality of the surgeon's movements in free space. Robotic surgical instrument vibrations are simple to measure and physically correspond to how roughly instruments are handled, but they have yet to be studied as a measure of technical surgical skill.
Methods:
Thirteen surgeons used a robotic surgery system (da Vinci S by Intuitive Surgical) to perform four trials each of peg transfer (PT), needle pass (NP), and intracorporeal suturing (IS). Completion time, instrument vibrations, and applied forces were measured for each trial; root mean square (RMS) and total sum of squares (TSS) were calculated from both the vibration and force recordings. Four experienced surgeons blindly assessed the task videos using a Global Rating Scale (GRS), and skill metrics were compared between the eight novices and five experienced participants. Stepwise regression was performed to predict GRS score from objective skill metrics. The concurrent validity of each metric was evaluated using receiver operating characteristic (ROC) analysis.
Results:
The GRS demonstrated excellent internal consistency (Cronbach's α = 0.91) and strong inter-rater reliability (ICC = 0.84). Compared to novices, experienced surgeons earned higher GRS scores and performed tasks with lower vibration magnitudes, lower forces, and shorter completion times in 15 of 18 task-metric combinations (p values ranging from 0.042 to <0.001). ROC analysis demonstrated that including vibration and force magnitudes along with completion time in skill prediction models improves the objective classification of subjects as novice or experienced for all tasks studied (PT: 90% sensitivity, 75% specificity; NP: 85% sensitivity, 84% specificity; suturing: 100% sensitivity, 100% specificity).
Conclusions:
RMS and TSS instrument vibrations are novel construct-valid measures of robotic surgical skill that enable the development of objective skill assessment models comparable to observer-based ratings.
Related Concept Videos
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of the Abdomen III: Palpation
Physical Assessment of the Respiratory Tract III: Percussion
Percussion in Respiratory Assessment
Percussion evaluates underlying tissue composition with audible and tactile vibrations,...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: