Related Experiment Video
Updated: Apr 6, 2026

Mass Spectrometry and Luminogenic-based Approaches to Characterize Phase I Metabolic Competency of In Vitro Cell Cultures
Published on: March 28, 2017
A selective and sensitive method based on UPLC-MS/MS for quantification of momordin Ic in rat plasma: application to
Huiyu Yan1, Yanqing Song1, Wei Zhou1
1Department of Pharmacy, the First Hospital of Jilin University, Changchun 130021, PR China.
Abstract:
A selective and sensitive method was developed and validated based on ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS). This method was applied to quantify momordin Ic in rat plasma. Chromatographic separation was performed on a Hypersil GOLD HPLC C18 column (150mm×4.6mm, 5μm) using an isocratic mobile phase of acetonitrile/water (80:20, v/v) at a flow rate of 0.6mL/min. An electrospray ionization source was applied and operated in negative ion mode; selected reaction monitoring (SRM) mode was used for quantification by monitoring the precursor-to-product ion transitions of m/z 763.4→m/z 455.3 for momordin Ic, and m/z 649.4→m/z 487.3 for IS. Calibration curves showed good linearity over the range of 22.0-2200ng/mL for momordin Ic in rat plasma. The developed method was applied to a pharmacokinetic study of momordin Ic in rats after single intravenous doses at 0.52, 1.56, and 4.67mg/kg. The elimination half-life (t1/2) values were 1.22±0.39, 1.14±0.10, and 1.83±0.39h, respectively. The plasma concentration at 2min (C2min) and area under the curve (AUC) for the intravenous doses of momordin Ic were approximately dose proportional.

