Validation of Numerically Simulated Tissue Temperatures During Transcutaneous Recharge of Neurostimulation Systems
John Patrick Abraham1, Brian Douglas Plourde1
1School of Engineering, University of St. Thomas, 2115 Summit Ave, St. Paul, MN, 55105-1079, USA.
Objective:
A research study combining numerical simulation and animal-model experiments has been performed to assess the ability of simulation to accurately calculate temperatures within living tissue during the recharge of a neuromodulation system (Restore Ultra device, Medtronic Neuromodulation, Minneapolis, MN, USA).
Materials And Methods:
The experiments were carried out on a sheep with the neuromodulation implant set to depths of 0.6 cm and 2.1 cm. Temperatures were recorded on the surfaces of the implant and on the sheep skin. Finite element simulations were carried out to determine the degree to which the simulations and experiments match. Additional calculations were performed for an intermediate implant depth.
Results:
It was found that there was excellent agreement between the simulations and the animal model for both depths.
Conclusion:
It is shown that numerical simulation using the Pennes bioheat equation is capable of predicting temperature increases within living tissues when implanted heat-generating devices are in use. The device used in the present study does not give rise to temperatures which cause concern of thermal injury or safety. The study was performed for aligned antenna and implant.
More Related Videos
08:50Fabrication and Characterization of a Conformal Skin-like Electronic System for Quantitative, Cutaneous Wound Management
Published on: September 2, 2015
07:52Author Spotlight: Advancing Human Brain Modulation – Optimized Protocols for Transcranial Ultrasound Stimulation Experiments
Published on: June 28, 2024
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...