Effect of display type, DICOM calibration and room illuminance in bitewing radiographs
Soili Kallio-Pulkkinen1,2, Sisko Huumonen3,4, Marianne Haapea1,2
11 Department of Diagnostic Radiology, Oulu University Hospital, Oulu, Finland.
Objectives:
To compare observer performance in the detection of both anatomical structures and caries in bitewing radiographs using consumer grade displays with and without digital imaging and communications in medicine (DICOM) calibration, tablets (third generation iPad; Apple, Cupertino, CA) and 6-megapixel (MP) displays under different lighting.
Methods:
30 bitewing radiographs were blindly evaluated on four displays under bright (510 lx) and dim (16 lx) ambient lighting by two observers. The dentinoenamel junction, enamel and dentinal caries, and the cortical border of the alveolar crests were evaluated. Consensus was considered as reference. Intraobserver agreement was determined. The proportion of equivalent ratings and weighted kappa were used to assess reliability.
Results:
The proportion of equivalent ratings with consensus differed significantly between uncalibrated and DICOM-calibrated consumer grade display in enamel caries in upper and lower molars in bright (p = 0.013 and p = 0.003) lighting, and in dentinal caries in lower molars in both bright (p = 0.022) and dim (p = 0.004) lighting. The proportion also differed significantly between DICOM-calibrated consumer grade and 6-MP display in dentinal caries in lower molars in bright lighting (p = 0.039), tablet and consumer grade display in enamel caries in upper molars (p = 0.017) in bright lighting, tablet and 6-MP display in dentinal caries in lower molars (p = 0.003) in bright lighting and in enamel caries in lower molars (p = 0.012) in dim lighting.
Conclusions:
DICOM calibration improves the detection of enamel and dentinal caries in bitewing radiographs, particularly in bright lighting. Therefore, a calibrated consumer grade display can be recommended as a diagnostic tool for viewing bitewing radiographs.
Related Concept Videos
X-ray Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...