Correcting systematic bias and instrument measurement drift with mzRefinery
Bryson C Gibbons1, Matthew C Chambers2, Matthew E Monroe1
1Biological Sciences Division, Pacific Northwest National Laboratory, Richland WA 99354 and.
Motivation:
Systematic bias in mass measurement adversely affects data quality and negates the advantages of high precision instruments.
Results:
We introduce the mzRefinery tool for calibration of mass spectrometry data files. Using confident peptide spectrum matches, three different calibration methods are explored and the optimal transform function is chosen. After calibration, systematic bias is removed and the mass measurement errors are centered at 0 ppm. Because it is part of the ProteoWizard package, mzRefinery can read and write a wide variety of file formats.
Availability And Implementation:
The mzRefinery tool is part of msConvert, available with the ProteoWizard open source package at http://proteowizard.sourceforge.net/
Contact:
samuel.payne@pnnl.gov.
Supplementary Information:
Supplementary data are available at Bioinformatics online.
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Random and Systematic Errors
Random and Systematic Errors
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Uncertainty in Measurement: Accuracy and Precision
NMR Spectrometers: Resolution and Error Correction